FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2083618 · Received May 10, 2011

Report

Report Number
2531779-2011-03313
Event Type
Malfunction
Date Received
May 10, 2011
Date of Event
April 10, 2011
Report Date
April 10, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 (B)(6) 2011 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: PUMP BOOTED TO 'VERIFY' SCREEN WITH FUNCTIONAL AUDITORY AND VIBRATORY ALARMS. THE PUMP BLACKBOX HISTORY SHOWS THAT THE PUMP REBOOTED MULTIPLE TIMES IN A CYCLE ON (B)(6) 2011. NO POWER LOSS ISSUES WERE REPRODUCED DURING INVESTIGATION. THE BATTERY CAP AND BATTERY COMPARTMENT WERE INTACT. THE BATTERY CAP ABLE TO SECURELY FASTEN TO THE PUMP WITH NO VISIBLE YELLOW O-RING.

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS FOR EVALUATION. THE EVALUATION OF THE PRODUCT HAS NOT BEEN COMPLETED; THEREFORE, NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. ONCE THE EVALUATION IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(6).

Description of Event or Problem · 1

ON (B)(6) 2011, THE REPORTER CONTACTED ANIMAS ON BEHALF OF HIS SON (THE PATIENT) ALLEGING THAT THE PUMP WAS REBOOTING. THE REPORTER REPORTED A BLOOD GLUCOSE LEVEL OF "285 MG/DL" BUT DENIED THAT THE PATIENT HAD ANY SYMPTOMS OF THIRST OR FREQUENT URINATION. THE PATIENT WAS NOT TESTED FOR KETONES. BASED ON THE INFORMATION PROVIDED, THIS COMPLAINT IS BEING REPORTED DUE TO THE FOLLOWING CONCLUSION: THE REPORTER CLAIMED THAT THE PUMP WAS REBOOTING. THERE IS NO EVIDENCE HOWEVER THAT THE ALLEGED ISSUE CONTRIBUTED TO A SERIOUS INJURY. THE REPORTER DID NOT REPORT ANY BLOOD GLUCOSE VALUES OR SYMPTOMS THAT MEET ANIMAS' CRITERIA FOR A SERIOUS INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 12 YR