GOLD RELOAD
Report
- Report Number
- 3005075853-2011-01911
- Event Type
- Malfunction
- Date Received
- May 10, 2011
- Date of Event
- April 13, 2011
- Report Date
- April 14, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K051002
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). CUSTOMER REPORTED THAT PACKAGE WAS BROKEN. ONE PACKAGE WAS RECEIVED FOR ANALYSIS WITH NO CARTON. PACKAGE WAS BRITTLE AND YELLOWED. YELLOWING OF THE PACKAGE SOMETIMES OCCURS DURING STERILIZATION PROCESS AND DOES NOT INDICATE ANY FAILURE OF PACKAGE INTEGRITY OR STERILIZATION. PACKAGE WAS VISUALLY EXAMINED FOR DAMAGE AND A RIP OR BREAK IN THE TYVEK WAS FOUND THAT IS FROM THE OUTSIDE IN AND IS ALIGNED WITH THE EDGE OF THE BLISTER FORM AND DEVICE. THE DAMAGE TO THE TYVEK APPEARS TO START AS A SCRAPE THAT DOES NOT BREAK THROUGH THE TYVEK AND APPEARS TO SCORE THE TYVEK IN A LINE THAT GOES DOWN THE LENGTH OF THE PACKAGE. IT APPEARS THAT COMPRESSION OF THE PACKAGE MAY HAVE CONTRIBUTED TO THE SEPARATION OF THE TYVEK ALONG THE SCORED LINE. THE RIP DOES NOT PENETRATE THE BLISTER FORM. DEFECT APPEARS TO BE AN ISOLATED INCIDENT, AS NO NONCONFORMANCES OR OTHER COMPLAINTS WERE ISSUED FOR SAME PROBLEM. COMPLAINT DATA AND NONCONFORMANCE DATA WAS REVIEWED FOR TIME PERIOD (B)(4), 2009 TO (B)(4), 2011. PACKAGING PROCESS WAS REVIEWED AND IT WAS CONCLUDED THAT NOTHING ON THE (B)(4) PROCESS COULD BE ESTABLISHED TO HAVE CAUSED THIS ISSUE. CAUSE AND TIME OF THE DAMAGE TO THE PACKAGE COULD NOT BE DETERMINED. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS DISCOVERED THAT THE STERILIZED PACKAGE OF THE PRODUCT WAS BROKEN BEFORE USE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GOLD RELOAD | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | UNK | G4UE0T |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |