FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2083543 · Received May 10, 2011

Report

Report Number
6000001-2011-03861
Event Type
Malfunction
Date Received
May 10, 2011
Date of Event
April 21, 2011
Report Date
April 21, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Removal / Correction Number
6000001-3/15/05-007-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 810:11 WAS CONFIRMED DURING PRODUCT EVALUATION IN THE PUMP'S EVENT HISTORY. THIS CONDITION WAS CAUSED BY THE AIR IN LINE PRINTED CIRCUIT BOARD (AIL PCB) BEING OUT OF CALIBRATION. THE AIL PCB WAS RECALIBRATED TO CORRECT THE REPORTED CONDITION. THIS ISSUE HAS BEEN ESCALATED TO CAPA. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION.

Description of Event or Problem · 1

DURING SERVICE BY BAXTER PERSONNEL, FAILURE CODE 810:11 WAS FOUND IN THE EVENT HISTORY TO HAVE CAUSED AN INTERRUPTION OF DELIVERY. THERE WAS NO KNOWN PATIENT INVOLVEMENT; THERE WERE NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS 6.13.90, CATEGORIZED AS REMEDIATED.