FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2083408
·
Received May 10, 2011
Report
- Report Number
- 2531779-2011-03299
- Event Type
- Malfunction
- Date Received
- May 10, 2011
- Date of Event
- February 18, 2011
- Report Date
- April 18, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION CONFIRMED THAT ALL KEYPAD BUTTONS ARE INTERMITTENTLY RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL BUTTON KEY CONTACTS.
Description of Event or Problem · 1
THE PATIENT REPORTED THAT THE KEYPAD BUTTONS HAVE BEEN SLOW TO RESPOND FOR THE (B)(6), AND NOW REQUIRE MULTIPLE HARD PRESSES TO OBTAIN A RESPONSE. SHE NOTED THAT THE BUTTONS INTERMITTENTLY CLICK AND SPRING BACK. THE PATIENT DENIED PHYSICAL DAMAGE TO THE RUBBER KEYPAD; DENIED EXPOSING THE PUMP TO WATER AND CLEANING PRODUCTS; AND STATED THAT SHE WEARS THE PUMP IN VARIOUS PLACES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR |