ACCU-CHEK SPIRIT COMBO
Report
- Report Number
- 2183996-2011-01281
- Event Type
- Injury
- Date Received
- April 28, 2011
- Date of Event
- April 5, 2011
- Report Date
- April 6, 2011
- Manufacturer
- ROCHE INSULIN DELIVERY SYSTEMS INC.
- Product Code
- LZG
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- UNKNOWN
Narratives
THIS INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED, IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.
ON (B)(6) 2011, PATIENT REPORTED SHE WAS ADMITTED TO THE HOSPITAL ON (B)(6) 2011, WITH HIGH KETONES AND VERY HIGH BLOOD GLUCOSE LEVELS. PATIENT STATED THAT INFUSION DEVICE HAD BEEN GIVING E6 (MECHANICAL ERROR) MESSAGES ALL DAY. PATIENT REPORTED THIS HAD BEEN OCCURRING FOR THE PAST 2 WEEKS. PATIENT STATED WHEN THE E6 ERRORS WERE OCCURRING; SHE HAD BEEN REPLACING THE BATTERY TO GET RID OF THE E6 ERRORS. PATIENT REPORTED SHE WAS GIVING BOLUSES ALL DAY BUT HER BLOOD GLUCOSE LEVEL WAS STILL READING "HI". PATIENT STATED SHE ALSO NOTICED INSULIN LEAKING INTO THE CARTRIDGE CHAMBER. PATIENT REPORTED SHE TRIED TO WASH IT OUT BUT IT CONTINUED TO HAPPEN. PATIENT STATED CRYSTALS FORMED AROUND THE PISTON ROD. PATIENT REPORTED SHE HAS NOW BEEN DISCHARGED FROM THE HOSPITAL AND IS WELL. PATIENT'S NORMAL BLOOD GLUCOSE LEVEL WAS NOT PROVIDED. PATIENT STATED SHE WAS PUT ON A SLIDING SCALE TO GET HER BLOOD GLUCOSE LEVELS DOWN. PATIENT REPORTED SHE WAS IN THE HOSPITAL FROM (B)(6) 2011. PATIENT STATED HER LAST BLOOD LEVEL PRIOR TO THE INCIDENT WAS BETWEEN 7-15 MMOL/L (126-270 MG/DL); EXACT READING WAS UNKNOWN. NO FURTHER INFORMATION IS AVAILABLE. PRODUCT WAS REPLACED AND REQUESTED RETURN OF THE ALLEGED PRODUCT FOR EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCU-CHEK SPIRIT COMBO | INSULIN INFUSION PUMP | LZG | ROCHE INSULIN DELIVERY SYSTEMS INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| O| R | INSULIN (DATE OF TX: (B)(6))| INSULIN INFUSION SET (DATE OF TX: (B)(6)) |