ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2083388 · Received May 10, 2011

Report

Report Number
2531779-2011-03297
Event Type
Malfunction
Date Received
May 10, 2011
Date of Event
April 18, 2011
Report Date
April 18, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 (B)(4) 2011-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS.

Additional Manufacturer Narrative · 1

(B)(6). THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

A FAMILY MEMBER REPORTED THAT THE KEYPAD BUTTONS BECAME UNRESPONSIVE AFTER THE PUMP WAS EXPOSED TO WATER. SHE NOTED THAT THE DISPLAY SCREEN SCROLLED WHILE CHANGING THE BATTERY, AND THEN THE BUTTONS BECAME UNRESPONSIVE. THE FAMILY MEMBER STATED THAT THE BUTTONS DO NOT CLICK AND SPRING BACK WHEN PRESSED. SHE DENIED PHYSICAL DAMAGE TO THE KEYPAD; DENIED EXPOSING THE PUMP TO CLEANING PRODUCTS; AND STATED THAT THE PATIENT WEARS THE PUMP IN A POUCH.