ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2011-03297
- Event Type
- Malfunction
- Date Received
- May 10, 2011
- Date of Event
- April 18, 2011
- Report Date
- April 18, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 1 (B)(4) 2011-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS.
(B)(6). THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
A FAMILY MEMBER REPORTED THAT THE KEYPAD BUTTONS BECAME UNRESPONSIVE AFTER THE PUMP WAS EXPOSED TO WATER. SHE NOTED THAT THE DISPLAY SCREEN SCROLLED WHILE CHANGING THE BATTERY, AND THEN THE BUTTONS BECAME UNRESPONSIVE. THE FAMILY MEMBER STATED THAT THE BUTTONS DO NOT CLICK AND SPRING BACK WHEN PRESSED. SHE DENIED PHYSICAL DAMAGE TO THE KEYPAD; DENIED EXPOSING THE PUMP TO CLEANING PRODUCTS; AND STATED THAT THE PATIENT WEARS THE PUMP IN A POUCH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 3 YR |