FDA Adverse Event Injury Summary report: N

M2A MAGNUM TAPER ADAPTER 42-50MM HEAD SIZE

MDR report key: 2083375 · Received May 10, 2011

Report

Report Number
1825034-2011-00365
Event Type
Injury
Date Received
May 10, 2011
Date of Event
March 28, 2011
Report Date
April 13, 2011
Manufacturer
BIOMET ORTHOPEDICS
Product Code
KWA
PMA / PMN Number
K042037
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

USER FACILITY FORWARDED A REPORT AFTER RECEIVING A FAXED LETTER FROM BIOMET ON (B)(6), 2011 WITH EVENT DETAILS. THIS FOLLOW-UP REPORT IS BEING FILED TO MAKE THE FDA AWARE THAT BOTH THE MANUFACTURER REPORT NUMBER AND USER FACILITY REPORT NUMBER REFERENCED IN THIS MEDWATCH ARE FOR THE SAME PATIENT, PART NUMBER AND EVENT. THIS REPORT FILED (B)(4), 2011.

Additional Manufacturer Narrative · 1

CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. EVALUATION OF THE RETURNED COMPONENT FOUND THAT IT WAS STILL ASSEMBLED WITH THE MODULAR HEAD AND NO ATTEMPT WAS MADE TO SEPARATE THE COMPONENTS. THE BACK SIDE OF THE TAPER ADAPTER SHOWED MINOR DAMAGE, POSSIBLY FROM SURGICAL REMOVAL. THE MORSE TAPER SHOWED EVIDENCE OF LIGHT WEAR AND DISCOLORATION FROM CONTACT WITH THE STEM. THE USER FACILITY WAS NOTIFIED OF THE EVENT ON (B)(6), 2011. TO DATE, A RESPONSE HAS NOT BEEN RECEIVED FROM THE USER FACILITY. IN THE EVENT THAT THE USER FACILITY SUBMITS A MEDWATCH REPORT, BIOMET WILL FORWARD A COPY TO THE FDA. THIS REPORT FILED (B)(4), 2011.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT UNDERWENT TOTAL HIP ARTHROPLASTY ON (B)(6), 2010. SUBSEQUENTLY, THE PATIENT BEGAN TO EXPERIENCE PAIN AND A REVISION PROCEDURE WAS PERFORMED ON (B)(6), 2011. THE ACETABULAR CUP, MODULAR HEAD AND TAPER ADAPTER WERE REMOVED AND REPLACED. THE STEM WAS WELL FIXED AND REMAINS IMPLANTED. NO FURTHER INFORMATION HAS BEEN PROVIDED TO DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 M2A MAGNUM TAPER ADAPTER 42-50MM HEAD SIZE PROSTHESIS, HIP KWA BIOMET ORTHOPEDICS N/A 510500

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R