FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 2083339
·
Received May 10, 2011
Report
- Report Number
- 2531779-2011-03291
- Event Type
- Malfunction
- Date Received
- May 10, 2011
- Date of Event
- April 18, 2011
- Report Date
- April 19, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. THE UP-ARROW KEYPAD BUTTON DID NOT ACTIVATE DESIRED PUMP FUNCTIONS DURING TESTING. THE UP-ARROW KEY CONTACT WAS FOUND TO BE INVERTED. THIS REPORT IS MADE UNDER THE REQUIREMENTS OF THE MEDICAL DEVICE REPORTING REGULATIONS AND DOES NOT CONSTITUTE AN ADMISSION ON THE PART OF ANIMAS OF ANY DEFICIENCY IN THE PERFORMANCE OF THE DEVICE.
Description of Event or Problem · 1
PER THE DISTRIBUTOR, IT WAS REPORTED THAT THE BUTTONS DO NOT WORK ANYMORE.