FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2083292 · Received May 10, 2011

Report

Report Number
2531779-2011-03286
Event Type
Malfunction
Date Received
May 10, 2011
Report Date
April 13, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED INTACT WITH NO LIFTING OR PEELING OBSERVED, THE PUMP WAS INTERMITTENTLY RESPONDING TO ALL KEYPAD BUTTONS, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE PATIENT CLAIMED THE KEYPAD WAS INTERMITTENTLY UNRESPONSIVE FOR PAST MONTH AND REPORTEDLY HAD TO PRESS BUTTONS SEVERAL TIMES FOR A RESPONSE. THE PATIENT REPORTED THAT KEYPAD IS INTACT AND THE BUTTONS FEEL NORMAL. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 23