FDA Adverse Event Malfunction Summary report: N

PARADIGM INSULIN INFUSION PUMP

MDR report key: 2083273 · Received April 13, 2011

Report

Report Number
3004209178-2011-81018
Event Type
Malfunction
Date Received
April 13, 2011
Date of Event
December 2, 2010
Report Date
April 7, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
LZG
PMA / PMN Number
K031390
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP FAILED THE DISPLACEMENT TEST DUE TO AN OUT-OF-PHASE MOTOR ENCODER SIGNAL. NO MOTOR ERROR ALARMS WERE NOTED DURING REWIND.

Description of Event or Problem · 1

IT WAS REPORTED THAT A MOTOR ERROR ALARM OCCURRED DURING THE MANUAL PRIME. THE ALARM HISTORY SHOWED MULTIPLE MOTOR ERROR ALARMS. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-712WWB

Patients

Seq Age Sex Outcome Treatment
1