FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2083248 · Received April 13, 2011

Report

Report Number
3004209178-2011-81047
Event Type
Malfunction
Date Received
April 13, 2011
Date of Event
March 25, 2011
Report Date
March 25, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP FAILED THE DISPLACEMENT TEST AND GAVE AN ALARM DURING REWIND DUE TO AN OUT-OF-PHASE MOTOR ENCODER SIGNAL. NO MOTOR ERROR ALARMS WERE NOTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A MOTOR ERROR ALARM HAD OCCURRED. THE CUSTOMER STATED THAT THEY WERE ABLE TO REWIND THE INSULIN PUMP AND PASS THE DISPLACEMENT TEST. THE CUSTOMER ALSO STATED THAT AN ALARM OCCURRED DURING A BOLUS AND ANOTHER ALARM OCCURRED MULTIPLE TIMES DURING REWIND. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAH

Patients

Seq Age Sex Outcome Treatment
1 15 YR