FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 2083248
·
Received April 13, 2011
Report
- Report Number
- 3004209178-2011-81047
- Event Type
- Malfunction
- Date Received
- April 13, 2011
- Date of Event
- March 25, 2011
- Report Date
- March 25, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP FAILED THE DISPLACEMENT TEST AND GAVE AN ALARM DURING REWIND DUE TO AN OUT-OF-PHASE MOTOR ENCODER SIGNAL. NO MOTOR ERROR ALARMS WERE NOTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT A MOTOR ERROR ALARM HAD OCCURRED. THE CUSTOMER STATED THAT THEY WERE ABLE TO REWIND THE INSULIN PUMP AND PASS THE DISPLACEMENT TEST. THE CUSTOMER ALSO STATED THAT AN ALARM OCCURRED DURING A BOLUS AND ANOTHER ALARM OCCURRED MULTIPLE TIMES DURING REWIND. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | MDS | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-722NAH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |