FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEMS

MDR report key: 2083192 · Received May 10, 2011

Report

Report Number
2050012-2011-01538
Event Type
Malfunction
Date Received
May 10, 2011
Date of Event
April 11, 2011
Report Date
April 11, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2011: THE FSE CHANGED ION SELECTIVE ELECTRODE (ISE) VALVES AND FOUND THE VACUUM WAS LOW IN WASTE TUBING 15 DUE TO A PINCH IN THE TUBING CLAMP. THE FSE REPLACED THE TUBING AND THE VACUUM WAS WITHIN SPECIFICATIONS. THE FSE PRIMED THE ISE AND EIC AND NO OVERFLOW WAS DETECTED. CALIBRATION AND QC WERE ACCEPTABLE. HARDWARE IS THE ROOT CAUSE FOR THIS EVENT.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) AND REPORTED THAT THEIR ELECTROLYTE INJECTION CUP (EIC) WAS OVERFLOWING ON UNICEL DXC 800 PRO SYNCHRON CLINICAL SYSTEM AND THEY HAD PROBLEM WITH QC. A CUSTOMER TECHNICAL SUPPORT (CTS) DIRECTED CUSTOMER TO DISASSEMBLE AND CLEAN EIC AND INSPECT VALVES PER IFU. CUSTOMER COMPLETED TROUBLESHOOTING RECOMMENDED BY CTS, BUT CONTINUED TO SEE OVERFLOWING DURING BUFFER PRIME. CTS GENERATED A SERVICE REQUEST FOR CUSTOMER. NO INJURY WAS REPORTED ON THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEMS CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. DXC 800 PRO N/A

Patients

Seq Age Sex Outcome Treatment
1