UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEMS
Report
- Report Number
- 2050012-2011-01538
- Event Type
- Malfunction
- Date Received
- May 10, 2011
- Date of Event
- April 11, 2011
- Report Date
- April 11, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
Narratives
A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2011: THE FSE CHANGED ION SELECTIVE ELECTRODE (ISE) VALVES AND FOUND THE VACUUM WAS LOW IN WASTE TUBING 15 DUE TO A PINCH IN THE TUBING CLAMP. THE FSE REPLACED THE TUBING AND THE VACUUM WAS WITHIN SPECIFICATIONS. THE FSE PRIMED THE ISE AND EIC AND NO OVERFLOW WAS DETECTED. CALIBRATION AND QC WERE ACCEPTABLE. HARDWARE IS THE ROOT CAUSE FOR THIS EVENT.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) AND REPORTED THAT THEIR ELECTROLYTE INJECTION CUP (EIC) WAS OVERFLOWING ON UNICEL DXC 800 PRO SYNCHRON CLINICAL SYSTEM AND THEY HAD PROBLEM WITH QC. A CUSTOMER TECHNICAL SUPPORT (CTS) DIRECTED CUSTOMER TO DISASSEMBLE AND CLEAN EIC AND INSPECT VALVES PER IFU. CUSTOMER COMPLETED TROUBLESHOOTING RECOMMENDED BY CTS, BUT CONTINUED TO SEE OVERFLOWING DURING BUFFER PRIME. CTS GENERATED A SERVICE REQUEST FOR CUSTOMER. NO INJURY WAS REPORTED ON THIS ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEMS | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 800 PRO | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |