FDA Adverse Event Injury Summary report: N

HAKIM INLINE PROGRAMMABLE VALVE SG

MDR report key: 2083014 · Received May 4, 2011

Report

Report Number
1226348-2011-00172
Event Type
Injury
Date Received
May 4, 2011
Manufacturer
CODMAN & SHURTLEFF, INC.
Product Code
JXG
PMA / PMN Number
K992173
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EXAMINATION OF THE VALVE REVEALED THE PRESENCE OF BIOLOGICAL DEBRIS WITHIN THE DEVICE. THIS BIOLOGICAL DEBRIS CAUSED THE RETURNED VALVE TO FAIL THE PROGRAMMING TEST AND IT APPEARS TO HAVE CAUSED THE DIFFICULTY ENCOUNTERED BY THE CUSTOMER. REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THE VALVE MET SPECIFICATION REQUIREMENTS WHEN RELEASED TO STOCK. BASED ON THE RESULTS OF THIS INVESTIGATION NO FURTHER ACTION IS REQUIRED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR COMPLAINTS. AT THE PRESENT TIME THIS COMPLAINT IS CLOSED.

Description of Event or Problem · 1

SALES REP REPORTED THAT APPROX FIVE YEARS AFTER IMPLANTATION THE VALVE WOULD NOT PROGRAM AND HAD TO BE REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HAKIM INLINE PROGRAMMABLE VALVE SG SHUNT, CENTRAL NERVOUS SYSTEM & COMPS JXG CODMAN & SHURTLEFF, INC. NA CHBBLH

Patients

Seq Age Sex Outcome Treatment
1 77 YR Required Intervention