FDA Adverse Event Injury Summary report: N

SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA

MDR report key: 2082843 · Received May 3, 2011

Report

Report Number
2023826-2011-00369
Event Type
Injury
Date Received
May 3, 2011
Date of Event
March 3, 2010
Report Date
April 6, 2011
Manufacturer
STAAR SURGICAL CO.
Product Code
HQL
PMA / PMN Number
P900048
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE PRODUCT PERTAINING TO THIS COMPLAINT WAS NOT RETURNED FOR EVALUATION. CONCLUSION: AN INVESTIGATION OF THE ROOT CAUSES OF LENS TEARS, SIMILAR TO THAT STATED IN THIS CLAIM, WERE ADDRESSED IN A CAPA OPENED IN (B)(4) 2005 (WHICH WAS SUBSEQUENTLY CLOSED). THE CAPA ADDRESSED DELIVERY SYSTEM ISSUES INCLUDING ALL STAGES OF MANUFACTURING OF THE INJECTORS AND CARTRIDGES. PROCESSES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. THE CAPA ALSO ADDRESSED HANDLING ERRORS. ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS. THE DAMAGE TO THE LENS COULD NOT BE EVALUATED BECAUSE LENS WAS NOT RETURNED TO STAAR. BASED ON THE COMPLAINT HISTORY, POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND HANDLING ERRORS BY THE CUSTOMER. (B)(4).

Description of Event or Problem · 1

THE REPORTER STATED THE SURGEON INSERTED AN (B)(4) THREE PIECE SILICONE LENS. THE LOOP HAPTIC TORE ON INSERTION INTO THE EYE. THE INCISION WAS ENLARGED TO REMOVE THE LENS, NO SUTURE USED. LENS WAS THROWN AWAY. BACKUP LENS WAS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA INTRAOCULAR LENS HQL STAAR SURGICAL CO. AQ2010V NA

Patients

Seq Age Sex Outcome Treatment
1 86 YR Required Intervention CARTRIDGE: MODEL AQ CARTRIDGE-FP, LOT NUMBER UNK| INJECTOR: MODEL MSI-TM, LOT NUMBER UNK