FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 2082674 · Received April 28, 2011

Report

Report Number
1811755-2011-01481
Event Type
Malfunction
Date Received
April 28, 2011
Date of Event
April 4, 2011
Report Date
April 4, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS EVALUATED BY THE MANUFACTURER AND THE COMPLAINT WAS DUPLICATED. THE TEMPERATURE RISE WAS FOUND TO BE DUE TO A DAMAGED SPINDLE HOUSING AND MOTOR CARTRIDGE, AND DUE TO ROTOR RUB. THEY WERE REPLACED AND THE UNIT WAS RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE OVERHEATED DURING A TOOTH EXTRACTION. THE PROCEDURE WAS COMPLETED WITH A BACK UP AND THERE WAS NO PROCEDURAL DELAY. THERE WERE NO ADVERSE CONSEQUENCES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CORE IMPACTION DRILL DZJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK