FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 2082674
·
Received April 28, 2011
Report
- Report Number
- 1811755-2011-01481
- Event Type
- Malfunction
- Date Received
- April 28, 2011
- Date of Event
- April 4, 2011
- Report Date
- April 4, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HI, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
DEVICE WAS EVALUATED BY THE MANUFACTURER AND THE COMPLAINT WAS DUPLICATED. THE TEMPERATURE RISE WAS FOUND TO BE DUE TO A DAMAGED SPINDLE HOUSING AND MOTOR CARTRIDGE, AND DUE TO ROTOR RUB. THEY WERE REPLACED AND THE UNIT WAS RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE OVERHEATED DURING A TOOTH EXTRACTION. THE PROCEDURE WAS COMPLETED WITH A BACK UP AND THERE WAS NO PROCEDURAL DELAY. THERE WERE NO ADVERSE CONSEQUENCES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CORE IMPACTION DRILL | DZJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |