FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 2082673 · Received April 28, 2011

Report

Report Number
1811755-2011-01454
Event Type
Malfunction
Date Received
April 28, 2011
Date of Event
March 30, 2011
Report Date
March 30, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

DURING QUALITY INVESTIGATION, THE CUSTOMER'S COMPLAINT WAS DUPLICATED. FURTHER INVESTIGATION REVEALED CORROSION DAMAGE TO THE MOTOR. ROTOR RUB, DAMAGED SPINDLE HOUSING AND MOTOR CARTRIDGE WERE IDENTIFIED AS THE PROBABLE CAUSE OF THE FAILURE. THE DAMAGED PARTS WERE REPLACED, PREVENTATIVE MAINTENANCE DONE AND THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

A STRYKER CORE IMPACTION DRILL WAS SENT IN FOR REPAIR DUE TO OVERHEATING DURING A PROCEDURE. NO ADVERSE EVENT WAS REPORTED; ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CORE IMPACTION DRILL DZJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK