FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 2082673
·
Received April 28, 2011
Report
- Report Number
- 1811755-2011-01454
- Event Type
- Malfunction
- Date Received
- April 28, 2011
- Date of Event
- March 30, 2011
- Report Date
- March 30, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
DURING QUALITY INVESTIGATION, THE CUSTOMER'S COMPLAINT WAS DUPLICATED. FURTHER INVESTIGATION REVEALED CORROSION DAMAGE TO THE MOTOR. ROTOR RUB, DAMAGED SPINDLE HOUSING AND MOTOR CARTRIDGE WERE IDENTIFIED AS THE PROBABLE CAUSE OF THE FAILURE. THE DAMAGED PARTS WERE REPLACED, PREVENTATIVE MAINTENANCE DONE AND THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
A STRYKER CORE IMPACTION DRILL WAS SENT IN FOR REPAIR DUE TO OVERHEATING DURING A PROCEDURE. NO ADVERSE EVENT WAS REPORTED; ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CORE IMPACTION DRILL | DZJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |