FDA Adverse Event
Malfunction
Summary report: N
MICRO RECIPROCATING SAW
MDR report key: 2082671
·
Received April 28, 2011
Report
- Report Number
- 1811755-2011-01453
- Event Type
- Malfunction
- Date Received
- April 28, 2011
- Date of Event
- March 29, 2011
- Report Date
- March 30, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
DURING QUALITY INVESTIGATION, THE CUSTOMER'S COMPLAINT WAS DUPLICATED. FURTHER INVESTIGATION REVEALED CORROSION DAMAGE TO THE MOTOR, ROTOR, BEARINGS, PRESS PLUG AND OTHER COMPONENT PARTS. THE DAMAGED PARTS WERE REPLACED AND THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
THE STRYKER MICRO RECIPROCATING SAW WAS SENT IN FOR REPAIR BECAUSE IT WAS OVERHEATING DURING A PROCEDURE. NO ADVERSE CONSEQUENCE WAS ASSOCIATED WITH THE EVENT; THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MICRO RECIPROCATING SAW | ERL | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |