FDA Adverse Event Malfunction Summary report: N

ADJ WIRE COLLET 0.7 - 1.8MM

MDR report key: 2082617 · Received April 27, 2011

Report

Report Number
1811755-2011-01462
Event Type
Malfunction
Date Received
April 27, 2011
Date of Event
April 6, 2011
Report Date
April 7, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
KIJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE MFR FOR INVESTIGATION. BASED ON THE QUALITY INVESTIGATION, THE IMPREGLON COATING ON THE DEVICE WAS WORN OFF. THIS CONDITION COULD CAUSE THE DEVICE TO STICK, AND NOT ALLOW THE COLLET TO RELEASE THE PIN.

Description of Event or Problem · 1

THE DEVICE WAS SENT IN FOR REPAIR WITH NO INFO. DURING THE REPAIR PROCESS, IT WAS DETERMINED THAT PINS WERE STICKING IN THE DEVICE. NO CONTACT INFO WAS INCLUDED WITH THE DEVICE, IT IS UNK IF THERE WAS ANY PT INVOLVEMENT OR ADVERSE CONSEQUENCES ASSOCIATED WITH THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADJ WIRE COLLET 0.7 - 1.8MM KIJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK