FDA Adverse Event
Malfunction
Summary report: N
ADJ WIRE COLLET 0.7 - 1.8MM
MDR report key: 2082617
·
Received April 27, 2011
Report
- Report Number
- 1811755-2011-01462
- Event Type
- Malfunction
- Date Received
- April 27, 2011
- Date of Event
- April 6, 2011
- Report Date
- April 7, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- KIJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO THE MFR FOR INVESTIGATION. BASED ON THE QUALITY INVESTIGATION, THE IMPREGLON COATING ON THE DEVICE WAS WORN OFF. THIS CONDITION COULD CAUSE THE DEVICE TO STICK, AND NOT ALLOW THE COLLET TO RELEASE THE PIN.
Description of Event or Problem · 1
THE DEVICE WAS SENT IN FOR REPAIR WITH NO INFO. DURING THE REPAIR PROCESS, IT WAS DETERMINED THAT PINS WERE STICKING IN THE DEVICE. NO CONTACT INFO WAS INCLUDED WITH THE DEVICE, IT IS UNK IF THERE WAS ANY PT INVOLVEMENT OR ADVERSE CONSEQUENCES ASSOCIATED WITH THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADJ WIRE COLLET 0.7 - 1.8MM | KIJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |