FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 2082553
·
Received April 28, 2011
Report
- Report Number
- 1811755-2011-01490
- Event Type
- Malfunction
- Date Received
- April 28, 2011
- Date of Event
- March 31, 2011
- Report Date
- March 31, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FURTHER INVESTIGATION REVEALED CORROSION DAMAGE TO THE MOTOR CARTRIDGE AND OTHER COMPONENT PARTS. ROTOR RUB, DAMAGED SPINDLE HOUSING AND CORRODED MOTOR CARTRIDGE WAS IDENTIFIED AS THE PROBABLE CAUSE OF THE FAILURE. THE DAMAGED PARTS WERE REPLACED; PREVENTIVE MAINTENANCE DONE AND THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
THE STRYKER CORE IMPACTION DRILL WAS SENT IN FOR REPAIR AND IT OVERHEATED DURING TESTING. NO ADVERSE CONSEQUENCE WAS ASSOCIATED WITH THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CORE IMPACTION DRILL | DZJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |