FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 2082553 · Received April 28, 2011

Report

Report Number
1811755-2011-01490
Event Type
Malfunction
Date Received
April 28, 2011
Date of Event
March 31, 2011
Report Date
March 31, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FURTHER INVESTIGATION REVEALED CORROSION DAMAGE TO THE MOTOR CARTRIDGE AND OTHER COMPONENT PARTS. ROTOR RUB, DAMAGED SPINDLE HOUSING AND CORRODED MOTOR CARTRIDGE WAS IDENTIFIED AS THE PROBABLE CAUSE OF THE FAILURE. THE DAMAGED PARTS WERE REPLACED; PREVENTIVE MAINTENANCE DONE AND THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

THE STRYKER CORE IMPACTION DRILL WAS SENT IN FOR REPAIR AND IT OVERHEATED DURING TESTING. NO ADVERSE CONSEQUENCE WAS ASSOCIATED WITH THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CORE IMPACTION DRILL DZJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK