FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 2082545 · Received April 28, 2011

Report

Report Number
1811755-2011-01480
Event Type
Malfunction
Date Received
April 28, 2011
Date of Event
April 4, 2011
Report Date
April 4, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS EVALUATED AT THE MANUFACTURER AND THE COMPLAINT OF SMOKE WAS DUPLICATED. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PRODUCT WAS GETTING AN OVERHEATING ERROR MESSAGE AND SMOKE CAME OUT THE DRILL DURING TESTING PRIOR TO A PROCEDURE. NO PT INVOLVEMENT OR ADVERSE CONSEQUENCES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CORE IMPACTION DRILL DZJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK