FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2082512
·
Received May 10, 2011
Report
- Report Number
- 2531779-2011-03261
- Event Type
- Malfunction
- Date Received
- May 10, 2011
- Date of Event
- March 14, 2011
- Report Date
- April 14, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION REVEALED THAT ALL KEYPAD BUTTONS ARE INTERMITTENTLY RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL BUTTON KEY CONTACTS.
Description of Event or Problem · 1
THE PATIENT REPORTED THAT THE OK KEYPAD BUTTON HAS BEEN STICKING FOR THE (B)(6). HE NOTED THAT THE BUTTON DOES NOT CLICK OR SPRING BACK, AND STATED THAT IT FEELS LIKE THERE IS GUM UNDERNEATH THE BUTTON. THE PATIENT REPORTED THAT THE KEYPAD IS WORN BUT INTACT; DENIED EXPOSING THE PUMP TO WATER AND CLEANING PRODUCTS; AND STATED THAT HE WEARS THE PUMP ON HIS BELT IN A CASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR |