FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2082512 · Received May 10, 2011

Report

Report Number
2531779-2011-03261
Event Type
Malfunction
Date Received
May 10, 2011
Date of Event
March 14, 2011
Report Date
April 14, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION REVEALED THAT ALL KEYPAD BUTTONS ARE INTERMITTENTLY RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL BUTTON KEY CONTACTS.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE OK KEYPAD BUTTON HAS BEEN STICKING FOR THE (B)(6). HE NOTED THAT THE BUTTON DOES NOT CLICK OR SPRING BACK, AND STATED THAT IT FEELS LIKE THERE IS GUM UNDERNEATH THE BUTTON. THE PATIENT REPORTED THAT THE KEYPAD IS WORN BUT INTACT; DENIED EXPOSING THE PUMP TO WATER AND CLEANING PRODUCTS; AND STATED THAT HE WEARS THE PUMP ON HIS BELT IN A CASE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 59 YR