FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2082472 · Received May 10, 2011

Report

Report Number
1423500-2011-05665
Event Type
Malfunction
Date Received
May 10, 2011
Date of Event
April 16, 2011
Report Date
April 16, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS REPORT FOR A SYSTEM ERROR 2240 (AIR IN SET) WAS NOT CONFIRMED; HOWEVER, PER THE COMPLAINT INFORMATION THE CAUSE OF THE SE 2240 IS DUE TO USE ERROR - AIR BEING SUCKED INTO THE DISPOSABLE AFTER THE PATIENT DISCONNECTED THE PATIENT LINE FROM THE TRANSFER SET. THERE WAS NO ALLEGATION REPORTED AGAINST THE BAXTER PRODUCT BY THE CUSTOMER; THEREFORE, THE SAMPLE WAS NOT REQUESTED FOR EVALUATION AND A BATCH REVIEW WILL NOT BE CONDUCTED. LABELING REVIEW FINDS THE LABELING ADEQUATE FOR THE USE ERROR IN THE COMPLAINT. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). DURING INVESTIGATION BY BAXTER THIS INCIDENT WAS DETERMINED TO BE CAUSED BY USE ERROR AND THERE WAS NO ALLEGATION OF A PRODUCT MALFUNCTION. THEREFORE, A BATCH REVIEW AND SAMPLE EVALUATION WILL NOT BE CONDUCTED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER AND REPORTED RECEIVING SYSTEM ERROR 2240 (AIR IN LINE) ALARM ON THE HOMECHOICE (HC) MACHINE DURING FILL. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE HOME PATIENT (HP). THE HP HAD DRAINED AND WAS GOING INTO A FILL. THE HP SAID SHE DID DISCONNECTED HERSELF IN HER THERAPY. THE TSR EXPLAINED THE ALARM AND HAD THE HP CYCLE POWER ON AND OFF. THE TSR EXPLAINED THE ALARM AND THE PROPER PROCEDURE FOR DISCONNECTING IN THERAPY. PRODUCT SURVEILLANCE CONTACTED THE NURSE ON (B)(6) 2011 REGARDING THE PATIENT'S USER ERROR OF DISCONNECTING DURING THERAPY. ACCORDING TO THE NURSE THE PATIENT RESUMED THERAPY AND REPORTED NO FURTHER ISSUES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 69 YR HOMECHOICE