FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 2082367 · Received May 9, 2011

Report

Report Number
2024168-2011-03318
Event Type
Injury
Date Received
May 9, 2011
Date of Event
March 1, 2011
Report Date
April 14, 2011
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE HAS BEEN RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: EVALUATION OF THE RETURNED DEVICE INDICATED THAT THE POSTERIOR CUFF MISS OCCURRED AS EVIDENCED BY THE POSTERIOR CUFF REMAINING IN THE FOOT POCKET. THE POSTERIOR NEEDLE TIP WAS RELEASED FROM THE SHANK AND ITS TIP WAS BENT, BUT THERE WAS NO NEEDLE STRIKE MARK OBSERVED AT THE POSTERIOR FOOT. THIS IS INDICATIVE OF THE POSTERIOR NEEDLE BEING DEFLECTED AWAY FROM THE POSTERIOR FOOT AND INTERACTED WITH HUMAN TISSUE INSTEAD OF ENGAGED WITH THE POSTERIOR CUFF INSIDE THE POSTERIOR FOOT POCKET DURING NEEDLE DEPLOYMENT AS DESIGNED. BECAUSE THE POSTERIOR NEEDLE DID NOT ENGAGE WITH THE POSTERIOR CUFF, THE CUFF WAS NOT EJECTED FROM THE FOOT. WHEN THE PLUNGER WAS REMOVED FROM THE DEVICE, THE LINK WAS HELD ON ONE END BY THE POSTERIOR CUFF IN THE FOOT POCKET WHILE BEING PULLED ON THE OTHER END BY THE WITHDRAWAL OF THE PLUNGER. THIS RESULTED IN THE ANTERIOR CUFF DETACHING FROM THE NEEDLE AS INDICATED BY DAMAGED ANTERIOR CUFF TABS. ONE OF THE ANTERIOR CUFF TABS WAS BROKEN OFF AND WAS NOT RETURNED WITH THE DEVICE. CUFF TABS MEASURE ONE ONE-HUNDREDTH (0.01) OF AN INCH SQUARE. DUE TO THE EXTREMELY SMALL SIZE, A MISSING CUFF TAB IS NOT VISIBLE WITHOUT MAGNIFICATION AN WOULD NOT BE NOTED BY THE USER. DURING TESTING, A PROXY PLUNGER WAS INSERTED TO TEST THE NEEDLE TRAJECTORY AND PUSH MANDREL TRAVEL LENGTH AND THE RESULTS MET THE MANUFACTURING CRITERIA. BASED ON THE INVESTIGATION FINDINGS, THE PROBABLE CAUSE FOR THE POSTERIOR CUFF MISS AND SUBSEQUENT ANTERIOR CUFF-TO-NEEDLE TIP DETACHMENT IS NEEDLE DEFLECTION DUE TO INTERACTION WITH HUMAN TISSUE OR A FAILURE TO MAINTAIN A STABLE POSITION OF THE DEVICE WITH RESPECT TO THE TISSUE TRACT. A REVIEW OF THE PRODUCT LOT HISTORY RECORD DID NOT REVEAL ANY NON-CONFORMING MATERIAL RECORDS ASSOCIATED WITH THIS LOT. NO MANUFACTURING OR QUALITY DEFICIENCY WAS DETECTED. THERE DOES NOT APPEAR TO BE ANY INDICATION OF A LOT SPECIFIC PRODUCT QUALITY DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE PROGLIDE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY AFTER AN UNSPECIFIED PROCEDURE. REPORTEDLY, A CUFF MISS OCCURRED. A SECOND PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WERE NO REPORTED ADVERSE PATIENT SEQUELAE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 010166H

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention