PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2011-03315
- Event Type
- Injury
- Date Received
- May 9, 2011
- Date of Event
- March 3, 2011
- Report Date
- April 14, 2011
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). EVALUATION SUMMARY: EVALUATION OF THE RETURNED DEVICE FOUND THE PLUNGER, SUTURE, AND NEEDLES WERE NOT RETURNED WITH THE DEVICE WHICH LIMITED THE SCOPE OF THE INVESTIGATION. INSPECTION OF THE RETURNED DEVICE INDICATED THAT THE POSTERIOR CUFF MISS OCCURRED AS EVIDENCED BY THE POSTERIOR CUFF REMAINING IN THE FOOT POCKET. THERE WAS A NEEDLE STRIKE MARK OBSERVED AT THE POSTERIOR FOOT, CONSISTENT WITH THE POSTERIOR NEEDLE STRIKING THE POSTERIOR FOOT DURING NEEDLE DEPLOYMENT INSTEAD OF ENGAGING WITH THE POSTERIOR CUFF INSIDE THE POSTERIOR FOOT POCKET AS DESIGNED. BECAUSE THE POSTERIOR NEEDLE DID NOT ENGAGE WITH THE POSTERIOR CUFF, THE CUFF WAS NOT EJECTED FROM THE FOOT. WHEN THE PLUNGER WAS REMOVED FROM THE DEVICE, THE LINK WAS HELD ON ONE END BY THE POSTERIOR CUFF IN THE FOOT POCKET WHILE BEING PULLED ON THE OTHER END BY THE WITHDRAWAL OF THE PLUNGER, THIS RESULTED IN THE ANTERIOR CUFF DETACHING FROM THE NEEDLE AS INDICATED BY BENT ANTERIOR CUFF TABS. DURING TESTING, A PROXY PLUNGER WAS INSERTED TO TEST THE NEEDLE TRAJECTORY AND PUSH MANDREL TRAVEL LENGTH AND THE RESULTS MET THE MANUFACTURING CRITERIA. BASED ON THE INVESTIGATION FINDINGS, THE PROBABLE CAUSE FOR THE POSTERIOR CUFF MISS AND SUBSEQUENT ANTERIOR CUFF-TO-NEEDLE TIP DETACHMENT IS NEEDLE DEFLECTION DUE TO INTERACTION WITH HUMAN TISSUE OR A FAILURE TO MAINTAIN A STABLE POSITION OF THE DEVICE WITH RESPECT TO THE TISSUE TRACT. NO MANUFACTURING OR QUALITY DEFICIENCY WAS DETECTED. A REVIEW OF THE PRODUCT LOT HISTORY RECORD DID NOT REVEAL ANY NON-CONFORMING MATERIAL RECORDS ASSOCIATED WITH THIS LOT. THERE DOES NOT APPEAR TO BE ANY INDICATION OF A LOT SPECIFIC PRODUCT QUALITY DEFICIENCY.
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE HAS BEEN RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.
IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE PROGLIDE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY AFTER AN UNSPECIFIED PROCEDURE. REPORTEDLY, A CUFF MISS OCCURRED. A SECOND PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WERE NO REPORTED ADVERSE PATIENT SEQUELAE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-TEMECULA-CT | 010406H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |