FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 2082344
·
Received May 9, 2011
Report
- Report Number
- 2531779-2011-03248
- Event Type
- Malfunction
- Date Received
- May 9, 2011
- Date of Event
- April 15, 2011
- Report Date
- April 15, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION AND EVALUATED. THE KEYPAD APPEARED TO BE INTACT; HOWEVER, ALL BUTTONS WERE INTERMITTENTLY RESPONDING TO USER INPUTS AND STICKING WHEN PRESSED DURING TESTING. WHEN THE KEYPAD WAS REMOVED IT WAS NOTED TO HAVE ADHESIVE UNDER THE CONTACTS. ALSO IT WAS NOTED THAT THE DEVICE WAS CRACKED AT BATTERY COMPARTMENT'S THREADS.
Description of Event or Problem · 1
THE PATIENT CLAIMED THAT THE UP/DOWN/OK BUTTONS REQUIRED SEVERAL PRESSES TO ACTIVATE THE DESIRED PUMP FUNCTIONS. THE KEYPAD IS REPORTEDLY INTACT AND HAS NOT BEEN EXPOSED TO MOISTURE. THERE WAS NO ADVERSE EVENT ASSOCIATE WITH THIS COMPLAINT. THE PUMP WAS REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR |