FDA Adverse Event Malfunction Summary report: N

COATED VICRYL PLUS ANTIBACTERIAL (POLYGLACTIN 910)

MDR report key: 2082276 · Received May 9, 2011

Report

Report Number
2210968-2011-00589
Event Type
Malfunction
Date Received
May 9, 2011
Date of Event
April 11, 2011
Report Date
April 13, 2011
Manufacturer
ETHICON, INC.
Product Code
GAM
PMA / PMN Number
K032420
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ACTUAL NEEDLE THAT BROKE AT THE POINT END REVEALED INDENTS AND SCUFF MARKS AROUND THE BREAK AREA PRODUCED DURING HANDLING BY THE SURGICAL NEEDLE HOLDERS OR SOME OTHER GRIPPING DEVICE. THE NEEDLE FRACTURED DUE TO TENSILE OVERLOAD GENERATED DURING SEVERE MECHANICAL DEFORMATION, WITH SIGNS OF DUCTILE FAILURE. THERE ARE NO SIGNS OF MANUFACTURING DEFECTS FOUND ON THE NEEDLE. THE DEVICE INFORMATION FOR USE CAUTIONS THE USER THAT "TO AVOID DAMAGING NEEDLE POINTS AND SWAGE AREAS, GRASP THE NEEDLE IN AN AREA ONE-THIRD (1/3) TO ONE-HALF (1/2) OF THE DISTANCE FROM THE SWAGED END TO THE POINT".

Additional Manufacturer Narrative · 1

DATE SENT TO THE FDA: (B)(4) 2011. (B)(4). CONCLUSION (B)(4): THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM. IN ADDITION, A REVIEW OF THE BATCH MANUFACTURING RECORDS WAS CONDUCTED AND THE BATCH MET ALL FINISHED GOODS RELEASE CRITERIA.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A STOMACH TUMOR RESECTION PROCEDURE ON (B)(6) 2011 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE BROKE WHEN THE SURGEON WAS HOLDING THE NEEDLE TIP WHILE SUTURING THE GASTROINTESTINAL ANASTOMOTIC STOMA. THE BROKEN PIECE DID NOT FALL INTO THE PATIENT. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COATED VICRYL PLUS ANTIBACTERIAL (POLYGLACTIN 910) SUTURE, ABSORBABLE GAM ETHICON, INC. NA CM2514

Patients

Seq Age Sex Outcome Treatment
1 62 YR