FDA Adverse Event Malfunction Summary report: N

ECHELON*FLEX60

MDR report key: 2082244 · Received May 9, 2011

Report

Report Number
1527736-2011-00110
Event Type
Malfunction
Date Received
May 9, 2011
Report Date
April 13, 2011
Manufacturer
ETHICON ENDO SURGERY, INC (CINCINNATI)
Product Code
GDW
PMA / PMN Number
K081146
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). ANVIL; JOINT COVER. THE ANALYSIS RESULTS FOUND THAT ONE EC60A DEVICE WAS RETURNED WITH THE ANVIL BENT UPWARDS AND WITHOUT CARTRIDGE RELOAD PRESENT. THE DAMAGE TO THE ANVIL IS CONSISTENT WITH THE DEVICE BEING CLAMPED OVER AN EXCESS OF TISSUE CAUSING THE ANVIL TO BENT AND FOR THE FIRING STROKE AND THE STAPLE FORM TO BE INCOMPLETE. IT SHOULD BE NOTED THAT A 100% INSPECTIONS TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRE SPECIFICATIONS PRIOR TO SHIPMENTS, IN ADDITION, A SAMPLE OF THE BATCH IS INSPECTED AT (B)(4). THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE STRAIGHT AND ARTICULATED POSITIONS WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THERE WERE MALFORMED STAPLES. THE INSTRUMENT COULD NOT BE CLOSED CORRECTLY, NO FURTHER INFORMATION IS AVAILABLE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ECHELON*FLEX60 STAPLE, IMPLANTABLE GDW ETHICON ENDO SURGERY, INC (CINCINNATI) UNK H4347J

Patients

Seq Age Sex Outcome Treatment
1