FDA Adverse Event Malfunction Summary report: N

COLLEAGUE P1.7

MDR report key: 2082239 · Received May 9, 2011

Report

Report Number
6000001-2011-03800
Event Type
Malfunction
Date Received
May 9, 2011
Date of Event
April 4, 2011
Report Date
April 5, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Removal / Correction Number
600001-3/15/05-007-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION WAS CONFIRMED BUT NOT DUPLICATED. THE ASSIGNABLE CAUSE WAS A FAULTY AIR IN LINE PRINTED CIRCUIT BOARD. THE AIR IN LINE PRINTED CIRCUIT BOARD HAS BEEN REPLACED. ADDITIONAL: THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS 6.13.90, CATEGORIZED AS REMEDIATED. A SERVICE HISTORY REVIEW HAS BEEN PERFORMED AND THE DEVICE WAS NOT PREVIOUSLY SERVICED FOR THE REPORTED CONDITION. A TREND REVIEW HAS BEEN PERFORMED AND THERE WERE SIMILAR REPORTS MADE TO BAXTER. THE ROOT CAUSE WILL CONTINUE TO BE INVESTIGATED THROUGH CAPA (CORRECTIVE AND PREVENTIVE ACTION) INVESTIGATION (B)(4).

Description of Event or Problem · 1

THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH A FAILURE CODE OF 810:11. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED, HOWEVER, IT IS KNOWN TO HAVE OCCURRED IN THE EMERGENCY ROOM. ACCORDING TO THE HOSPITAL REPRESENTATIVE, THERE WAS NO PATIENT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION REPORTED RELATED TO THE DEVICE. DURING THE PRODUCT EVALUATION AT BAXTER, IT WAS DISCOVERED THAT FAILURE CODE 810:11 OCCURRED DURING INFUSION, WHICH CAUSED AN INTERRUPTION DURING DELIVERY. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS CURRENTLY UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE P1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1