FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 2082200
·
Received May 9, 2011
Report
- Report Number
- 2531779-2011-03238
- Event Type
- Malfunction
- Date Received
- May 9, 2011
- Date of Event
- April 13, 2011
- Report Date
- April 13, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. ALL KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS DURING TESTING. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.
Description of Event or Problem · 1
THE PATIENT REPORTED THAT OVER THE PAST TWO MONTHS THE KEYPAD IS INTERMITTENTLY RESPONDING. THE PATIENT DENIES ANY TEARS OR PEELING OF THE KEYPAD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR |