FDA Adverse Event Malfunction Summary report: N

UNIVERSAL HIGH TORQUE DRILL

MDR report key: 2082110 · Received April 26, 2011

Report

Report Number
1811755-2011-01438
Event Type
Malfunction
Date Received
April 26, 2011
Date of Event
March 21, 2011
Report Date
April 1, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DRILL WAS RECEIVED BY THE U.S. MANUFACTURER AND THE COMPLAINT WAS CONFIRMED. INTERNAL COMPONENTS WERE EVALUATED AND A WORN MOTOR ASSEMBLY AND BEARING WAS DISCOVERED. A DAMAGED MOTOR ASSEMBLY AND BEARING CAN LEAD TO INCREASED FRICTION BETWEEN ROTATING COMPONENTS, RESULTING IN HEAT BEING GENERATED. ADDITIONALLY, A CUT WAS DISCOVERED IN THE POWER CORD, WHICH CAN ALSO CAUSE THE MOTOR ASSEMBLY TO GENERATE HEAT WHEN THE DRILL IS OPERATED. THE MOTOR ASSEMBLY AND BEARING WERE REPLACED, AND THE DEVICE WAS RETURNED TO THE USER FACILITY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING CONDUCTED BY A MANUFACTURER TECHNICAL SPECIALIST, THE DRILL HEATED UP. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, SO THERE WAS NO PATIENT INVOLVEMENT. NO USER INJURY WAS REPORTED, AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNIVERSAL HIGH TORQUE DRILL ERL STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK