FDA Adverse Event
Malfunction
Summary report: N
ADJUSTABLE PIN COLLET
MDR report key: 2082099
·
Received April 26, 2011
Report
- Report Number
- 1811755-2011-01430
- Event Type
- Malfunction
- Date Received
- April 26, 2011
- Date of Event
- March 31, 2011
- Report Date
- March 31, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- GET
- PMA / PMN Number
- K943323
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR INVESTIGATION. BASED ON THE QUALITY INVESTIGATION, THE IMPREGLON COATING ON THE DEVICE WAS WORN OFF. THIS CONDITION COULD CAUSE THE DEVICE TO STICK, AND NOT ALLOW THE COLLET TO RELEASE THE PIN.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WOULD NOT RELEASE THE PIN DURING ROUTINE MAINTENANCE. THERE WAS NO PATIENT INVOLVEMENT AND NO ALLEGATION OF ADVERSE CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADJUSTABLE PIN COLLET | GET | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |