FDA Adverse Event Malfunction Summary report: N

ANGLED LONG SABER ATTACHMENT

MDR report key: 2082088 · Received April 26, 2011

Report

Report Number
1811755-2011-01420
Event Type
Malfunction
Date Received
April 26, 2011
Date of Event
March 29, 2011
Report Date
March 29, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DRILL ATTACHMENT WAS RETURNED TO THE MANUFACTURER AND THE COMPLAINT WAS CONFIRMED. BASED ON THE RESULTS OF THE QUALITY INVESTIGATION, THE BEARINGS WERE HEATING UP DUE TO CONTAMINATION AND DEBRIS. DEBRIS WITHIN BEARINGS WILL CAUSE FRICTION, WHICH CAN LEAD TO HEAT BEING GENERATED. IMPROPER OR INADEQUATE CLEANING/STERILIZATION TECHNIQUES COULD CONTRIBUTE TO CORROSION OF INTERNAL COMPONENTS WITHIN THIS DEVICE. THE DRILL ATTACHMENT COULD NOT BE REPAIRED AND THEREFORE WAS NOT RETURNED TO THE USER FACILITY.

Description of Event or Problem · 1

IT WAS REPORTED THAT A DRILL ATTACHMENT WAS LOUD AND HEATED UP DURING A ROUTINE MAINTENANCE EVALUATION BY THE MANUFACTURER'S FIELD SERVICE TEAM. THIS EVENT DID NOT OCCUR IN A SURGICAL ENVIRONMENT. THERE WAS NO USER INJURY REPORTED AND NO ADVERSE CONSEQUENCES ALLEGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANGLED LONG SABER ATTACHMENT ERL STRYKER INSTRUMENTS KALAMAZOO 10302

Patients

Seq Age Sex Outcome Treatment
1 UNK