FDA Adverse Event
Malfunction
Summary report: N
ADJ PIN COLLET 2.0 - 3.2MM
MDR report key: 2082078
·
Received April 26, 2011
Report
- Report Number
- 1811755-2011-01404
- Event Type
- Malfunction
- Date Received
- April 26, 2011
- Date of Event
- April 1, 2011
- Report Date
- April 1, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- KIJ
- PMA / PMN Number
- K972367
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO THE MFR FOR INVESTIGATION. BASED ON THE QUALITY INVESTIGATION, THE IMPREGLON COATING ON THE DEVICE WAS WORN OFF. THIS CONDITION COULD CAUSE THE DEVICE TO STICK, AND NOT ALLOW THE COLLET TO RELEASE THE PIN.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PIN WAS STICKING IN THE DEVICE DURING A PROCEDURE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH THIS DEVICE WITH NO DELAY OR ALLEGATION OF ADVERSE CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADJ PIN COLLET 2.0 - 3.2MM | KIJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |