FDA Adverse Event Malfunction Summary report: N

MAESTRO DRILL

MDR report key: 2082056 · Received April 26, 2011

Report

Report Number
1811755-2011-01419
Event Type
Malfunction
Date Received
April 26, 2011
Date of Event
March 28, 2011
Report Date
March 30, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
HBB
PMA / PMN Number
K041754
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED BY THE MFR; HOWEVER THE LEAKING AIR COMPLAINT COULD NOT BE REPLICATED. THE DRILL MOTOR HOSE ASSEMBLY WAS REPLACED, AND PREVENTATIVE MAINTENANCE WAS PERFORMED. THE DEVICE WAS REPAIRED AND RETURNED TO THE USER FACILITY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING, THE PNEUMATIC DRILL WAS LEAKING AIR. THERE WAS NO REPORT OR ANY OIL LEAKAGE FROM THE DEVICE. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, SO THERE WAS NO PT INVOLVEMENT. NO USER INJURY WAS REPORTED, AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAESTRO DRILL HBB STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK