SOVEREIGN COMPACT
Report
- Report Number
- 2020664-2011-00022
- Event Type
- Malfunction
- Date Received
- April 26, 2011
- Date of Event
- March 30, 2011
- Report Date
- March 30, 2011
- Manufacturer
- ABBOTT MEDICAL OPTICS (AMO)
- Product Code
- HQC
- PMA / PMN Number
- K003638
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
THE PHACOEMULSIFICATION MACHINE WAS EXAMINED AND TESTED BY AN AMO SVC TECHNICIAN AT THE CUSTOMER'S LOCATION. THE TECHNICIAN INDICATED THE UNIT WOULD NOT START UP AND RESEATED THE SUBSYSTEM CONTROLLER BOARD RAM CHIP. THE ISSUE WAS RESOLVED. THE TECHNICIAN CALIBRATED VACUUM SETTINGS AND THE UNIT PASSED PRIME AND TUNE PROCESS. THE SYS WAS VERIFIED FOR ALL MODE OF OPERATIONS AND CALIBRATIONS. THE SYS HAS CONTINUED TO BE USED WITHOUT ANY FURTHER REPORTED INCIDENTS. THE MACHINE OPERATES TO AMO SPECIFICATIONS. NO ADD'L INFO WAS PROVIDED.
THE CLINIC REPORTED AT THE END OF COMPLETION OF A CATARACT PROCEDURE, THE CLINIC WAS IN IRRIGATION/ASPIRATION MODE AND THE PHACOEMULSIFICATION MACHINE SHUT DOWN ON ITS OWN. IN ADDITION, THE CLINIC INDICATED THE MACHINE SHUT DOWN DURING PROCEDURE ONE WEEK PRIOR TO REPORTING. AT THE TIME, THE CLINIC WAS ABLE TO RESTART THE SAME MACHINE AND COMPLETE CASE. THE SURGEON COMPLETED THE PROCEDURES WITH A GOOD SURGICAL OUTCOME. NO PT INJURY REPORTED.