FDA Adverse Event Malfunction Summary report: N

SOVEREIGN COMPACT

MDR report key: 2082055 · Received April 26, 2011

Report

Report Number
2020664-2011-00022
Event Type
Malfunction
Date Received
April 26, 2011
Date of Event
March 30, 2011
Report Date
March 30, 2011
Manufacturer
ABBOTT MEDICAL OPTICS (AMO)
Product Code
HQC
PMA / PMN Number
K003638
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PHACOEMULSIFICATION MACHINE WAS EXAMINED AND TESTED BY AN AMO SVC TECHNICIAN AT THE CUSTOMER'S LOCATION. THE TECHNICIAN INDICATED THE UNIT WOULD NOT START UP AND RESEATED THE SUBSYSTEM CONTROLLER BOARD RAM CHIP. THE ISSUE WAS RESOLVED. THE TECHNICIAN CALIBRATED VACUUM SETTINGS AND THE UNIT PASSED PRIME AND TUNE PROCESS. THE SYS WAS VERIFIED FOR ALL MODE OF OPERATIONS AND CALIBRATIONS. THE SYS HAS CONTINUED TO BE USED WITHOUT ANY FURTHER REPORTED INCIDENTS. THE MACHINE OPERATES TO AMO SPECIFICATIONS. NO ADD'L INFO WAS PROVIDED.

Description of Event or Problem · 1

THE CLINIC REPORTED AT THE END OF COMPLETION OF A CATARACT PROCEDURE, THE CLINIC WAS IN IRRIGATION/ASPIRATION MODE AND THE PHACOEMULSIFICATION MACHINE SHUT DOWN ON ITS OWN. IN ADDITION, THE CLINIC INDICATED THE MACHINE SHUT DOWN DURING PROCEDURE ONE WEEK PRIOR TO REPORTING. AT THE TIME, THE CLINIC WAS ABLE TO RESTART THE SAME MACHINE AND COMPLETE CASE. THE SURGEON COMPLETED THE PROCEDURES WITH A GOOD SURGICAL OUTCOME. NO PT INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SOVEREIGN COMPACT HQC ABBOTT MEDICAL OPTICS (AMO) CMP680300

Patients

Seq Age Sex Outcome Treatment
1