FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2082045 · Received April 26, 2011

Report

Report Number
3004209178-2011-81224
Event Type
Malfunction
Date Received
April 26, 2011
Date of Event
April 14, 2011
Report Date
April 14, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER'S BLOOD GLUCOSE WAS DROPPING RAPIDLY AND UNEXPECTEDLY. THE CUSTOMER STATED THAT HER BLOOD GLUCOSE READING DROPPED FROM 456 MG/DL TO 311 MG/DL WITHIN HALF AN HOUR AFTER SHE BOLUSED WITH THE BOLUS WIZARD. PROGRAMMING WAS CORRECT. TROUBLESHOOTING WAS PERFORMED AND THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-522NAP

Patients

Seq Age Sex Outcome Treatment
1 39 YR