FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2081985 · Received May 9, 2011

Report

Report Number
1423500-2011-05635
Event Type
Malfunction
Date Received
May 9, 2011
Date of Event
April 14, 2011
Report Date
April 14, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING INVESTIGATION BY BAXTER, THIS INCIDENT WAS DETERMINED TO BE CAUSED BY USE/USER ERROR AND THERE WAS NO ALLEGATION OF A PRODUCT MALFUNCTION. THEREFORE A BATCH REVIEW AND SAMPLE EVALUATION WILL NOT BE CONDUCTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW UP MDR WILL BE SENT.

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT WAS CONFIRMED BASED ON INFORMATION PROVIDED BY THE PATIENT. LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE IN REGARD TO THIS ISSUE. THERE IS NOT ENOUGH INFORMATION TO DETERMINE THE CAUSE OF THE USE ERROR, THEREFORE THE ROOT CAUSE IS UNDETERMINED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

DURING TROUBLESHOOTING FOR A CHECK LINES AND BAGS ALARM THAT OCCURRED ON THE HOMECHOICE (HC) DURING DWELL 5. THE HOME PATIENT (HP) STATED ONLY SUPPLY BAG HAS SOLUTION AND HEATER AND FINAL BAGS ARE EMPTY. THE HP ALSO STATED THAT EARLY ON HE HAD MISTAKENLY CONNECTED FINAL BAG TO SUPPLY LINE AND SUPPLY BAG TO FINAL LINE- HE HAD THEN SWITCHED THE BAGS BACK OVER TO CORRECT LINES. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) HAD THE HP CYCLE POWER ON MACHINE TO END OF THERAPY AND THEN GO TO MANUAL DRAIN TO DRAIN OUT. THE TSR EXPLAINED THE ALARM AND POSSIBLE CAUSES AND ASSISTED THE HP TO END THERAPY AND DRAIN MANUALLY. THERE WAS PATIENT INVOLVEMENT AND THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT.