FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2081952 · Received April 5, 2011

Report

Report Number
1720753-2011-03252
Event Type
Malfunction
Date Received
April 5, 2011
Date of Event
March 18, 2011
Report Date
April 5, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE MANUFACTURER'S SERVICE REPRESENTATIVE PERFORMED AN ON-SITE INVESTIGATION. THE BATTERY PACK WAS REPLACED AND THE FILAMENT WAS CALIBRATED. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED. THIS EVENT COULD BE A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 9800 SYSTEM DISPLAYED A HIGH MILLIAMP ERROR MESSAGE. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1