FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2081853 · Received May 9, 2011

Report

Report Number
2531779-2011-03232
Event Type
Malfunction
Date Received
May 9, 2011
Date of Event
April 13, 2011
Report Date
April 13, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 (B)(6) 2011 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. UNRELATED TO THIS COMPLAINT, THE CARTRIDGE CAP WAS REPORTED TO HAVE A TEAR. (B)(6).

Description of Event or Problem · 1

THE PATIENT'S MOTHER REPORTED THAT (B)(6) AGO THE KEYPAD BUTTONS STARTED TO STICK AND NOT RESPOND TO PRESSES. THE MOTHER REPORTED THERE IS NO DAMAGE TO THE KEYPAD AND NO EXPOSURE TO WATER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1 9 YR