ANIMAS INSULIN INFUSION PUMP
MDR report key: 2081811
·
Received May 9, 2011
Report
- Report Number
- 2531779-2011-03229
- Event Type
- Malfunction
- Date Received
- May 9, 2011
- Date of Event
- April 14, 2011
- Report Date
- April 14, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION REVEALED THAT ALL KEYPAD BUTTONS ARE INTERMITTENTLY RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL BUTTON KEY CONTACTS.
Description of Event or Problem · 1
THE PATIENT REPORTED THAT THE OK KEYPAD BUTTON REQUIRES SEVERAL PRESSES TO OBTAIN A RESPONSE. SHE NOTED THAT THE OK BUTTON FEELS DEPRESSED AND DOES NOT CLICK. THE PATIENT CONFIRMED THAT THE KEYPAD IS INTACT; AND DENIED EXPOSING THE PUMP TO WATER AND CLEANING PRODUCTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR |