FDA Adverse Event Malfunction Summary report: N

ALINITY I PROCESSING MODULE

MDR report key: 20817987 · Received December 2, 2024

Report

Report Number
3016438761-2024-00688
Event Type
Malfunction
Date Received
December 2, 2024
Date of Event
November 22, 2024
Report Date
January 27, 2025
Manufacturer
ABBOTT LABORATORIES
Product Code
JJE
UDI-DI
00380740137366
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FIELD SERVICE REPRESENTATIVE (FSR) INSPECTED THE INSTRUMENT AND OBSERVED THAT THE ALNTY C PWR CN EU WAS CHARRED. THE PART WAS REPLACED WHICH RESOLVED THE ISSUE. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. AN INSTRUMENT SERVICE HISTORY REVIEW REVEALED NO ADDITIONAL ISSUES OF CHARRING RELATED TO THE ALNTY C PWR CN EU REPORTED FOR (B)(6). A REVIEW OF TRACKING AND TRENDING OF THE ALINITY I DID NOT IDENTIFY ANY TRENDS ASSOCIATED WITH THE COMPLAINT ISSUE. A REVIEW OF TRACKING AND TRENDING FOR THE ALNTY C PWR CN EU DID NOT IDENTIFY ANY TRENDS. A REVIEW OF THE MANUFACTURING DOCUMENTATION DID NOT IDENTIFY ANY ISSUES ASSOCIATED WITH THE COMPLAINT ISSUE. LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE. THE 2025 UL CERTIFICATION MEMO INDICATES THAT ABBOTT DIAGNOSTIC EQUIPMENT AND ACCESSORIES ARE CERTIFIED TO THE APPROPRIATE SAFETY STANDARDS, AND ADEQUATE PROTECTION IS PROVIDED FOR THE OPERATOR AGAINST SPREAD OF FIRE FROM THE EQUIPMENT. THE CHARRING OBSERVED WAS LIMITED TO ALNTY C PWR CN EU; THE CHARRING DID NOT SPREAD TO OTHER PARTS OF THE MODULE. BASED ON THE AVAILABLE INFORMATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ALINITY I PROCESSING MODULE FOR SERIAL (B)(6) OR THE ALNTY C PWR CN EU WAS IDENTIFIED.

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Description of Event or Problem · 0

THE CUSTOMER REPORTED THE ALINITY I PROCESSING MODULE STOPPED RUNNING. UPON INSPECTION, THE CUSTOMER FOUND THE ELECTRICAL OUTLET WAS OXIDIZED AND HEARD A SIZZLING SOUND. THE INSTRUMENT WAS UNPLUGGED, AND THE PLUG WAS HOT TO THE TOUCH. UPON INSPECTION THE ELECTRICAL CONNECTOR OF THE POWER CABLE WAS FOUND TO CHARRED. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED. NO HARM TO THE USER WAS REPORTED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED THE ALINITY I PROCESSING MODULE STOPPED RUNNING. UPON INSPECTION, THE CUSTOMER FOUND THE ELECTRICAL OUTLET WAS OXIDIZED AND HEARD A SIZZLING SOUND. THE INSTRUMENT WAS UNPLUGGED, AND THE PLUG WAS HOT TO THE TOUCH. UPON INSPECTION THE ELECTRICAL CONNECTOR OF THE POWER CABLE WAS FOUND TO CHARRED. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED. NO HARM TO THE USER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
93778 ALINITY I PROCESSING MODULE ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE ABBOTT LABORATORIES 00380740137366

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown