FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2081786 · Received May 9, 2011

Report

Report Number
1423500-2011-05622
Event Type
Malfunction
Date Received
May 9, 2011
Date of Event
April 14, 2011
Report Date
April 15, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS REPORT FOR THE SYSTEM ERROR 2240 (AIR IN LINE) WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE. NOT ENOUGH DATA IS AVAILABLE WITHIN THE COMPLAINT TO IDENTIFY ROOT CAUSE; THEREFORE, THE ROOT CAUSE OF THE COMPLAINT WAS NOT DETERMINED. THE LOT NUMBER IS UNKNOWN THEREFORE A BATCH REVIEW WAS NOT PERFORMED. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE IS NOT AVAILABLE FOR EVALUATION, THEREFORE, BAXTER CANNOT DETERMINE THE ROOT CAUSE. SINCE THE LOT NUMBER IS UNKNOWN A BATCH REVIEW WILL NOT BE PERFORMED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER AND REPORTED RECEIVING SYSTEM ERROR 2240 (AIR IN LINE) ALARM ON THE HOMECHOICE (HC) MACHINE DURING DWELL 2 OF 3. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE HOME PATIENT (HP) EXPLAINED THE ALARM. THE TSR ADVISED THE CAREGIVER (CG) TO START OVER WITH NEW SUPPLIES. PRODUCT SURVEILLANCE CONTACTED THE HP ON (B)(6) 2011. THE HP STATED THEY STARTED THERAPY WITH ALL NEW SUPPLIES. THE HP STATED THAT THE ISSUE WAS RESOLVED; HOWEVER, THE CAUSE OF THE ALARM REMAINED UNKNOWN. THE HP VERIFIED THAT THERE WERE NO LOOSE CONNECTIONS, DISCONNECTIONS OR DEFECTS ON THE SUPPLIES. THE HP STATED THAT THEY WERE DOING FINE AND CONTINUING THERAPY WITHOUT ISSUES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 66 YR HOMECHOICE