FDA Adverse Event
Malfunction
Summary report: N
SIZE #4 4-IN-1 CUTTING BLOCK CAPTURED ASSY. TRIA.
MDR report key: 2081779
·
Received April 14, 2011
Report
- Report Number
- 2249697-2011-00475
- Event Type
- Malfunction
- Date Received
- April 14, 2011
- Date of Event
- March 22, 2011
- Report Date
- March 22, 2011
- Manufacturer
- STRYKER ORTHOPAEDICS MAHWAH
- Product Code
- LXH
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
EVAL SUMMARY: A MFG NON-CONFORMANCE WAS OBSERVED BASED ON VISUAL INSPECTION WHICH INDICATED THAT THE FIXATION PEG(S) DISASSOCIATED FROM THE POSTERIOR FACE OF THE TRIATHLON CUTTING BLOCK. THE PRODUCT EXPERIENCE REPORTING DATABASE SHOWS THIS EVENT IS RELATED TO A TREND OF SIMILAR EVENTS WHERE THE FIXATION PEG(S) HAS DISASSOCIATED FROM THE POSTERIOR FACE OF THE TRIATHLON CUTTING BLOCK.
Description of Event or Problem · 1
IT WAS REPORTED: "SMALL PIN ON BACK OF CUTTING BLOCK, BROKE OFF BLOCK AS IT WAS BEING EXTRACTED FROM PT'S KNEE. THE PIN WAS REMOVED."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SIZE #4 4-IN-1 CUTTING BLOCK CAPTURED ASSY. TRIA. | INSTRUMENT | LXH | STRYKER ORTHOPAEDICS MAHWAH | NA | SB3A06 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Other |