FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 2081744 · Received April 11, 2011

Report

Report Number
3002158293-2011-00404
Event Type
Malfunction
Date Received
April 11, 2011
Date of Event
March 4, 2011
Report Date
April 8, 2011
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHECK BELT MESSAGES) HAS BEEN CONFIRMED. UPON EVALUATION, IT WAS FOUND THAT A PULSE WIRE ON ECG B WAS NOT CONNECTED. THE CAUSE FOR THE DISCONNECTED PULSE WIRES CANNOT BE POSITIVELY DETERMINED BUT WAS LIKELY THE RESULT OF COLD SOLDERS. NO ADVERSE EVENT RESULTED FROM THE OPEN CONNECTION. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) MALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT HE IS RECEIVING CONSTANT CHECK BELT ALARMS. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 79 YR