LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3002158293-2011-00396
- Event Type
- Malfunction
- Date Received
- April 11, 2011
- Date of Event
- February 28, 2011
- Report Date
- April 8, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY PACK SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (BATTERY WON'T HOLD CHARGE) HAS BEEN CONFIRMED. AS RECEIVED, THE BATTERY WAS FOUND TO HAVE A POORLY SOLDERED THERMISTOR. THE THERMISTOR WAS NOT SOLDERED CORRECTLY TO THE PAD ON THE BATTERY BOARD PCA. THIS PREVENTED THE BATTERY FROM PROPERLY CHARGING ON THE BATTERY CHARGER. THE ROOT CAUSE OF THE POOR SOLDER CONNECTION CANNOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY AN ASSEMBLY ERROR. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY PACK. THE PT RECEIVED A REPLACEMENT BATTERY PACK.
A (B)(6) FEMALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT HER CHARGER WOULD NOT RECOGNIZE HER BATTERIES. THE PT WAS ISSUED A REPLACEMENT BATTERY CHARGER AND BATTERY PACKS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 38 YR |