FDA Adverse Event Malfunction Summary report: N

M SERIES

MDR report key: 2081716 · Received April 12, 2011

Report

Report Number
1220908-2011-00925
Event Type
Malfunction
Date Received
April 12, 2011
Report Date
March 22, 2011
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K990762
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORP HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A F/U REPORT WHEN OUR INVESTIGATION IS COMPLETED.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S DISPLAY WAS DAMAGED AND WAS NOT LEGIBLE. COMPLAINANT INDICATED THAT THERE WAS NO PT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 M SERIES DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION MSERIES BI-PHASIC NA

Patients

Seq Age Sex Outcome Treatment
1 NA