ALINITY C CREATININE REAGENT KIT
Report
- Report Number
- 3002809144-2024-00339
- Event Type
- Malfunction
- Date Received
- December 2, 2024
- Date of Event
- October 31, 2024
- Report Date
- January 17, 2025
- Manufacturer
- ABBOTT GMBH
- Product Code
- CGX
- UDI-DI
- 00380740173470
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A REVIEW OF TICKETS DETERMINED THAT THERE IS NORMAL COMPLAINT ACTIVITY FOR ALINITY CREATININE LN 7P99 LOT 79666UN24. TRENDING REVIEW DETERMINED NO TRENDS FOR FALSELY ELEVATED RESULTS FOR THE PRODUCT. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. A REVIEW OF THE MANUFACTURING DOCUMENTATION DID NOT IDENTIFY ANY ISSUES ASSOCIATED WITH THE CUSTOMER'S OBSERVATION. A REVIEW OF THE PRODUCT LABELING CONCLUDED THAT THE ISSUE IS SUFFICIENTLY ADDRESSED. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ALINITY CREATININE LN 7P99 LOT 79666UN24 WAS IDENTIFIED.
AN EVALUATION IS IN PROCESS. A FOLLOW UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.
THE CUSTOMER REPORTED A FALSELY ELEVATED CREATININE RESULT GENERATED ON THE ALINITY C PROCESSING MODULE ON A 74-YR OLD FEMALE PATIENT. RESULTS PROVIDED: (B)(6) 2024 SID (B)(6) = 333.77 UMOL/L, NEW SAMPLE ON ANOTHER ALINITY. (B)(6) 2024 SID (B)(6) = 59.04 UMOL/L (REFERENCE RANGE: 45.00-133.00 UMOL/L). RE-TESTED BOTH SAMPLES ON (B)(6) 2024 ON DIFFERENT ALINITY: (B)(6) 2024 SID (B)(6) = 352.38 UMOL/L. (B)(6) 2024 SID (B)(6) = 60.8 UMOL/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
THE CUSTOMER REPORTED A FALSELY ELEVATED CREATININE RESULT GENERATED ON THE ALINITY C PROCESSING MODULE ON A 74-YR OLD FEMALE PATIENT. RESULTS PROVIDED: ON (B)(6) 2024, SID: (B)(6); 333.77 UMOL/L, NEW SAMPLE ON ANOTHER ALINITY ON (B)(6) 2024, SID: (B)(6); 59.04 UMOL/L (REFERENCE RANGE: 45.00-133.00 UMOL/L). RE-TESTED BOTH SAMPLES ON (B)(6) 2024 ON DIFFERENT ALINITY: ON (B)(6) 2024, SID: (B)(6); 352.38 UMOL/L, ON (B)(6) 2024, SID: (B)(6); 60.8 UMOL/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 327704 | ALINITY C CREATININE REAGENT KIT | ALKALINE PICRATE, COLORIMETRY, CREATININE | CGX | ABBOTT GMBH | 79666UN24 | 00380740173470 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR | Female | ALNTY C PROCESSING MODU, 03R67-01, AC05399| ALNTY C PROCESSING MODU, 03R67-01, AC05399 |