FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEMS

MDR report key: 2081677 · Received May 9, 2011

Report

Report Number
2050012-2011-01534
Event Type
Malfunction
Date Received
May 9, 2011
Date of Event
April 10, 2011
Report Date
April 10, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED AND REPLACED THE NO FOAM BOTTLE ASSEMBLY. NO FURTHER LEAKS OBSERVED BY FSE. HARDWARE IS THE ROOT CAUSE FOR THIS EVENT.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC., (BCI) AND REPORTED A LEAK OF NO FOAM FROM A CRACKED TUBE OR CONNECTOR TO NO FOAM BOTTLE ON UNICEL DXC 800 PRO SYNCHRON CLINICAL SYSTEM. THE CUSTOMER WAS UNABLE TO IDENTIFY EXACT LOCATION OF THE LEAK. NO INJURY WAS REPORTED ON THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEMS CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. DXC 800 PRO N/A