FDA Adverse Event
Malfunction
Summary report: N
ACCU-CHEK SPIRIT
MDR report key: 2081622
·
Received April 13, 2011
Report
- Report Number
- 2183996-2011-01006
- Event Type
- Malfunction
- Date Received
- April 13, 2011
- Date of Event
- February 4, 2011
- Report Date
- March 15, 2011
- Manufacturer
- ROCHE INSULIN DELIVERY SYSTEMS INC.
- Product Code
- LZG
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- UNKNOWN
Narratives
Description of Event or Problem · 1
THE PT REPORTED, THE UP BUTTON OF THE INFUSION DEVICE IS DEFECTIVE. NO PHYSIOLOGICAL EFFECTS WERE REPORTED. THE INFUSION DEVICE WAS REPLACED AND REQUESTED TO BE RETURNED FOR EVAL. UPON EVAL OF THE INFUSION DEVICE, IT WAS FOUND THE UP/DOWN BUTTON IS ALWAYS ACTIVE. DUE TO AN EXTERNAL MECHANICAL INFLUENCE, THE SNAP DOME OF THE UP/DOWN BUTTON IS PRESSED THROUGH. THIS SOURCE LED TO AN ALWAYS ACTIVATED UP/DOWN BUTTON AND, THEREFORE, ALL THE BUTTONS DO NOT REACT ON PRESSURE. THE ISSUE IS RELATED TO USER HANDLING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCU-CHEK SPIRIT | INSULIN INFUSION PUMP | LZG | ROCHE INSULIN DELIVERY SYSTEMS INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | INSULIN INFUSION SET| INSULIN |