FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2081602 · Received April 13, 2011

Report

Report Number
2531779-2011-02528
Event Type
Malfunction
Date Received
April 13, 2011
Report Date
March 18, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. ALL KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS DURING TESTING. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS. IT WAS FOUND THAT THE SYMBOLS WERE WORN OFF THE KEYPAD.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE UP, DOWN, AND OK BUTTONS ARE INCREASINGLY UNRESPONSIVE TO PRESSES. IT WAS REPORTED THAT THE PUMP IS NOT EXPOSED TO WATER AND THERE IS NO KEYPAD PEELING. IT WAS REPORTED THAT THE SYMBOLS ARE WORN OFF THE KEYPAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1